The amount of mineral and salt impurities in the water is called total dissolved solids (TDS). TDS shows how many units of impurities there are for one million units of water. One way to measure impurities in water is to measure the electric conductivity of water. Absolutely pure water is a poor conductor of electricity. Water shows significant conductivity when dissolved salts are present. Over most ranges, the amount of conductivity is directly proportional to the amount of salts dissolved in the water or Salinity. Four basic styles of Conductivity / TDS / Salinity meters are available for varying applications: Pens, Portable Meters, Data Loggers, and Benchtop meters.
Rugged housing with large LCD displaying the parameter being
measured together with time, date and temperature (in C or F)
Automatic or manual temperature compensation
Automatic or manual ranging
Automatic buffer recognition
Digital and analog outputs
Min/Max and 99 data points
Data Hold
5 calibration points on Conductivity and TDS (total dissolved solids) and 2 for Salinity
Electrode status indication
Operates on AC voltage (no batteries). Comes ready to use with ATC conductivity electrode, electrode holder arm, AC adaptor, computer cable and software.
Dimensions: 8 x 7 x 2 (216 x 178 x 57 mm)
Dimensions with Holder Arm: 10 x 8 x 9 (254 x 203 x 229 mm)
Weight: 18.7 oz (530 g)
| Range | Resolution | Accuracy | |
| Conductivity uS/cm | 0 ~ 19.9 0 ~ 199.9 0 ~ 1999 | 0.01 0.1 1 | 1% fs + 1d |
| Conductivity mS/cm | 0 ~ 19.99 0 ~ 199.9 | 0.01 0.1 | |
| TDS ppm | 0 ~ 19.99 0 ~ 199.9 0 ~ 1999 | 0.01 0.1 1 | |
| TDS ppt | 0 ~ 19.99 0 ~ 199.9 | 0.01 0.1 | |
| Salinity ppt | 0 ~ 11.38 0 ~ 80 | 0.01 0.1 | 1% fs + 1 digit |
| Temperature F | 23 ~ 176 | 0.1 | 0.9F |
| Temperature C | -5 ~ 80 | 0.1 | 0.5C |